                  12W Medical Power Adapter With IEC/EN60601 Standards, Suitable for Medical Equipments ▼
                    product:12W Medical Power Adapter With IEC/EN60601 Standards, Suitable for Medical Equipments
                      KRECO P/N:KRE012MPS Series
                      Datasheets Download
                      Catalogs Download
                      Inquiry:​ sales@kreco.com.cn
                      Explore Power Ecosystem: www.ipskre.com
                Details:
                 ▶12W Medical Use AC/DC Adaptor Series Wall-Mount
                12V 700mA medical use AC/DC adaptor 12W series wall-mount, Universal input, customized design
                Applications for: medical equipment, health device and more

                ▶Features
                High-reliability
                LED indication
                Light and convenient plug-in
                High-efficiency: consumption, meet US Energy Start/CEC/AS/MEPS/EUP/ERP
                Suit medical equipment, health device

                ▶Typical model list
                Input Condition DC Output
                Voltage
                DC Output
                Current
                Output Voltage
                Precision
                Ripple & Noises Regulation
                Input
                Voltage:
                90V-264Vac
                Frequency:
                47-63Hz
                1 5V 2.0A -5%-+8% 50mV Line Load
                2 5.5V 2.0A  ±5% 50mV ±1% ±5%
                3 7V 1.4A  ±5% 70mV ±1% ±5%
                4 7V 1.6A  ±5% 70mV ±1% ±5%
                5 9V 1.3A  ±5% 90mV ±1% ±5%
                6 12V 0.7A  ±5% 120mV ±1% ±5%
                7 12V 1.0A  ±5% 120mV ±1% ±5%
                8 15V 0.8A  ±5% 150mV ±1% ±5%
                9 24V 0.5A  ±5% 240mV ±1% ±5%
                10 - - - - - -

                ▶Protections
                Thermal shutdown capability
                Over-voltage protection
                Over-current protection
                Short-circuit protection

                ▶Reliability
                Burn-in: 100% full load, 40°C, 4 hours (minimum)
                MTBF: meet MIL-HDBK-217F over 50,000 hours, full load, 25°C
                Leakage current: 0.25mA maximum at 264V AC
                Withstanding voltage: 3kV, 1 minute, 10mA (maximum), between primary and secondary

                ▶Emission and immunity
                EN60601
                EN55011

                ▶Working environmental
                Storage humidity: 10 to 95% RH
                Operating humidity: 10 to 95% RH
                Storage temperature: -20 to 65°C
                Operating temperature:  0 to 40°C

                ▶Safety approvals
                CB(TUV-SUD)+LVD:IEC/EN6060